Description
The LD-3 Dielectic Cell is designed for testing rigid flat materials. Material up to 8cm or 3 1/8 inches with a maximum thickness of 1.1 cm or 7/16 inches can be used in the dielectric cell. The dielectric cell is made up of two electrodes. One electrode is fixed with a grounded guard ring and the other is movable via a micrometer. The micrometer allows precise measurement of electrode spacing. The grounded guard ring reduces fringe effects to improve measurements.
The dielectric cell has a usable frequency range from 50 Hz to 1MHz.
The three terminal electrical connections are made via GR-874 Connection on the fixed electrode side, optional BNC to GR-874 adapter is available, and banana connection on the movable electrode side.
A 1 Meter BNC to BNC-T cable set is available for connection to the cell.
Specifications
Size: 4” x 4” x 7.5” depth
Weight: 8.1 lbs
Materials: 303 Stainless and Teflon
Terminals: Three Terminals with grounded guard ring
Connectors: GR874 and banana jack
Electrodes: 2.5” fixed measuring electrode with 2.87 OD guard ring, 2.87” movable electrode
Sample Size: 0.005” to 0.35” thick and 3.125” square or 2.87” diameter
Frequency Range: 50Hz to 1MHz
Options