Medical SystemsSoftwareHipot TestersLCR MetersMegohmmetersMilliohmmetersCable TestersGround Bond TestersAccessoriesRefurbished EquipmentProductos de QuadTech
Additional Images: Front Rear Right
The Guardian 12KV DC performs DC Hipot testing over a voltage range of 100 to 12,000V DC with a maximum leakage current output of 10mA. Additionally, the 12KV tester provides insulation resistance measurement capability over a range 100kΩ to 50GΩ, at test voltages from 50 to 1000V DC.
In all cases, both high and low limits are operator programmable for quick, and easy pass/fail testing. These dual limits are essential to ensure that the device under test is actually making contact with the tester. For consistent testing, up to 50 setups can be stored in instrument memory for later recall. An optional high voltage scanner is also available for testing multi-connection devices.
The following accessories come standard with the unit:
The following accessory options are available: